Table 1.

Intra- and inter-day precision and accuracy for HCY and CYS

Concentration (μM)
% RSD % Bias
Spiked Measured (mean ± SD)
Intra-assay reproducibilitya
Quality Controls
 HCY 4.0 3.8 ± 0.1 2.9 −5.5
16.0 16.2 ± 0.4 2.5 1.4
80.0 80.5 ± 1.6 2.0 0.6
 CYS 32.5 31.4 ± 0.4 1.4 −3.2
130.0 132.5 ± 3.2 2.4 1.9
650.0 645.3 ± 13.3 2.1 −0.7
Inter-assay reproducibilityb
Quality Controls
 HCY 4.0 3.8 ± 0.3 7.9 −4.5
16.0 16.1 ± 0.7 4.3 0.9
80.0 81.6 ± 3.4 4.2 2.0
 CYS 32.5 31.8 ± 1.9 5.8 −2.3
130.0 131.7 ± 6.0 4.6 1.3
650.0 657.8 ± 31.5 4.8 1.2
Standards
 HCY 3.13 3.2 ± 0.1 3.3 0.7
6.25 6.4 ± 0.2 3.2 2.1
12.5 13.0 ± 0.2 1.7 3.9
25.0 22.9 ± 0.5 2.4 −8.3
50.0 50.9 ± 0.2 0.5 1.9
100.0 100.7 ± 0.6 0.6 0.7
 CYS 25.0 25.8 ± 0.7 2.5 3.2
50.0 50.0 ± 1.1 2.3 0.0
100.0 100.5 ± 1.1 1.1 0.5
200.0 186.0 ± 4.7 2.5 −7.0
400.0 406.0 ± 4.2 1.0 1.5
800.0 807.8 ± 5.5 0.7 1.0
a

Twelve quality control samples per concentration.

b

Six to twelve quality control samples or two standards per day per concentration for three days.