Published in final edited form as: J Chromatogr B Analyt Technol Biomed Life Sci. 2007 Feb 21;852(1-2):554–561. doi: 10.1016/j.jchromb.2007.02.024

Table 2.

Intra- and inter-day precision and accuracy for CysGly and GSH

Concentration (μM)
% RSD % Bias
Spiked Measured (mean ± SD)
Intra-assay reproducibilitya
Quality Controls
 CysGly 7.5 7.3 ± 0.1 1.5 −3.3
30.0 29.8 ± 0.7 2.4 −0.7
150.0 164.2 ± 3.1 1.9 9.5
 GSH 2.0 2.1 ± 0.1 4.1 7.4
8.0 8.1 ± 0.2 2.1 1.5
40.0 38.9 ± 0.7 1.7 −2.9
Inter-assay reproducibilityb
Quality Controls
 CysGly 7.5 7.4 ± 0.4 5.1 −1.7
30.0 29.9 ± 1.1 3.6 −0.3
150.0 165.0 ± 5.4 3.3 10
 GSH 2.0 2.1 ± 0.1 4.9 5.5
8.0 8.2 ± 0.3 3.6 2.8
40.0 39.1 ± 1.2 3.1 −2.3
Standards
 CysGly 6.25 6.3 ± 0.2 2.7 0.6
12.5 12.5 ± 0.2 1.9 0.1
25.0 25.2 ± 0.4 1.5 0.8
50.0 48.3 ± 1.0 2.2 −3.4
100.0 102.3 ± 0.3 0.3 2.3
200.0 199.3 ± 1.0 0.5 −0.4
 GSH 1.56 1.5 ± 0.1 4.6 −4.4
3.13 3.4 ± 0.1 3.0 7.4
6.25 6.5 ± 0.1 2.1 4.0
12.5 12.3 ± 0.2 1.8 −1.7
25.0 24.5 ± 0.3 1.1 −1.9
50.0 50.4 ± 0.2 0.4 0.7
a

Twelve quality control samples per concentration.

b

Six to twelve quality control samples or two standards per day per concentration for three days.