Published in final edited form as: Anal Biochem. 2006 May 11;355(1):71–80. doi: 10.1016/j.ab.2006.04.045

Table 4.

Precision and accuracy data and Z′ values acquired from the fluorescent endpoint system validation

Nominal inhibition (%) Day 1
Day 2
Day 3
Interday precisionb (CV %)
Mean accuracya (%) Precisiona (CV %) Mean accuracya (%) Precisiona (CV %) Mean accuracya (%) Precisiona (CV %)
Blank 3.4 2.2 3.4 4.6
Control 4.6 7.3 6.7 6.2
Inhibitors
CEU 5 95.2 5.2 71.3 3.9 59.0 3.5 5.0
CHU 35 97.9 3.7 92.4 2.0 70.0 6.2 5.7
CDU 55 99.9 2.5 95.0 3.9 88.6 2.6 5.7
AUDA 65 98.9 3.5 98.9 2.2 87.1 2.5 6.0
DCU 95 98.6 2.8 99.8 3.0 98.3 2.9 4.9
Z 0.8c 0.8c 0.7c 0.7d

Note. Mean accuracy was determined by calculating the ratio of the determined inhibition percentage divided by the nominal value and multiplying it by 100%. Precision or coefficient of variation (CV) values represent standard deviations of fluorescent signal divided by the mean fluorescent signal of maximum enzyme activity and multiplied by 100%.

a

Intraday (n = 16).

b

Intraday (n = 48).

c

Intraday average (n = 4).

d

Interday average (n = 12).