Skip to main content
. 2007 Jun 15;3(4):387–392.

Table 3.

Sensitivity and Specificity Analysis: ApneaLink(home) vs ApneaLink (lab)

AHI Sensitivity
Specificity
PPV
NPV
AUC
2 h 4 h 2 h 4 h 2 h 4 h 2 h 4 h 2 h 4 h
≥ 5 90.0 90.0 75.0 66.7 92.3 90.0 69.2 66.7 0.881 0.878
≥ 10 83.3 81.3 78.6 73.9 76.9 68.4 84.6 85.0 0.874 0.842
≥ 15 76.2 84.6 93.6 92.3 88.9 84.6 85.3 92.3 0.922 0.923
≥ 20 76.5 81.8 94.3 96.4 86.7 90.0 89.2 93.1 0.931 0.971

The apnea-hypopnea index (AHI) obtained from ApneaLink study in the home versus the ApneaLink AHI obtained during the sleep laboratory study. 2 hours (at-home and in-lab) = 52 subjects. 4 hours (at-home and in-lab) = 39 subjects. PPV refers to positive predictive value; NPV, negative predictive value; AUC, area under the curve.