Table 1.

Testing guidelines recommended by ISO, USP and ASTM for an implantable device that will contact the body tissue or bone for more than 30 days *.

Regulatory Body Recommended Standards
ISO Initial Evaluation Tests
Cytotoxicity
ISO 10993-5: Tests for in-vitro cytotoxicity
Sensitization
ISO 10993-10: Tests for irritation and delay-type hypersensitivity
Genotoxicity
ISO 10993-3: Tests for genotoxicity, carcinogenicity and reproductive toxicity
Implantation
ISO 10993-6: Tests for local effects after implantation

Supplementary Evaluation Tests
Chronic Toxicity
ISO 10993-11: Tests for Systemic Toxicity
Carcinogenicity
ISO 10993-3: Tests for genotoxicity, carcinogenicity and reproductive toxicity
USP Initial Evaluation Tests
Cytotoxicity
Biological Reactivity Tests, In Vitro <87>
Sensitization
Procedure to come
Genotoxicity
Procedure to come
Implantation
Biological Reactivity Tests, In Vivo <88>
Chronic toxicity
Procedure to come
Carcinogenicity
Procedure to come

Additional Tests that may be applicable
Irritation or Intracutaneous Reactivity
Biological Reactivity Tests, In Vivo <88>
Systemic Toxicity (Acute)
Biological Reactivity Tests, In Vivo <88>
Subchronic Toxicity (Subacute toxicity)
Procedure to come
ASTM Cell Culture Cytotoxicity
Practice F813 and Test Method F895
Application specific method in Practice F 1027
Sensitization
Practice F 720
Repeated dermal patch test method also acceptable
Skin Irritation or Intracutaneous
Practice F 719 (skin irritation)
Practice F749 (intracutaneous injection) or USP <88>
Systemic Toxicity, Acute or Subchronic
Practice F750 or USP <88>
Pyrogen Test
USP <85> Bacterial Endotoxins Test
Short-term Implantation
Practice F 1408 – short term subcutaneous
Practice F 763 – short term intramuscular or USP <88>
Long-term Implantation
Practice F 981 other long-term implant tests may also be appropriate
Immune Response (Immunotoxicity)
No ASTM standard provided
Genotoxicity
Ames test (preliminary screen of materials)
Guide E1262 and Guide E1280
Warns that no single test can detect all possible mutagens
Carcinogenicity
No standard procedures available
Guide F 1439
*

Table 1 based on references [4850] for ISO, ASTM and USP, respectively.