Fig. 4.
a) Axial LSF and b) relative range lobe levels for no paddle and lexan 375 μm paddle with beamforming corrections. Range lobe levels, which include all reverberations, were 9 dB greater when imaging through the lexan paddle than when imaging through the TPX paddle or no paddle, primarily due to the acoustical impedance mismatch between lexan and saline.