Table 2.
Resulting jet disrupter voltages and integrated signal intensities after jet disrupter regulation of the ion flux to match the ion current of the 10 pmol/μL sample.
Sample Conc. (pmol/μL) | Jet disrupter Voltage (V) | Ion Current (pA) | Integrated Signal Intensity | ISI variation from 10 pmol/μL sample (%) |
---|---|---|---|---|
10 | 178 | 480 | 4.54E+07 | -- |
15 | 185 | 480 | 4.57E+07 | 0.7 |
20 | 190 | 480 | 4.47E+07 | −1.5 |
25 | 193 | 480 | 4.60E+07 | 1.3 |