Skip to main content
. Author manuscript; available in PMC: 2007 Sep 27.
Published in final edited form as: J Acoust Soc Am. 2006 Aug;120(2):676–685. doi: 10.1121/1.2214131

TABLE III.

Acoustic intensity, power output, and electric-to-acoustic energy conversion efficiency of the 3.3 MHz HIFU transducer.

Output Vpp (V) ISPPA (W/cm2) ISAPA (W/cm2) ISAPA/ISPPA Pin (W) P (W) P/Pin
0.05 240 127 0.53 1.10 0.72 65.4%
0.1 981 551 0.56 4.38 2.92 66.8%
0.2 3728 2338 0.63 17.52 12.71 72.5%
0.3 7088 4801 0.68 40.75 28.05 68.8%
0.4 9730 6849 0.70 73.36 50.54 68.9%
0.5 10 967 8144 0.74 109.56 62.90 57.4%