Table 2 Summary of results from participating laboratories for the UK National External Quality Assessment Scheme (NEQAS) pilot HER2 fluorescence in situ hybridisation (FISH) scheme.
Results | Appropriate | Acceptable | Inappropriate | Misdiagnosis | |
---|---|---|---|---|---|
SK‐BR‐3 | |||||
Run 1 | 22 (9) | 10 | 8 | 4 | 0 |
Run 2 | 27 (5) | 9 | 10 | 8 | 0 |
Run 3 | 29 (2) | 16 | 3 | 10 (1NR) | 1 |
Total | 78 (16) | 35 | 21 | 20 | 1 |
MDA‐MB‐453 | |||||
Run 1 | 22 (9) | 13 | 3 | 6 | 2* (<1.8–2.0) |
Run 2 | 27 (5) | 17 | 3 | 7 | 6 |
Run 3 | 29 (2) | 15 | 6 | 8 | 3 |
Total | 78 (16) | 45 | 12 | 21 | 11 |
BT20/MDA‐MB‐175* | |||||
Run 1 | 22 (9) | 16 | 2 | 4 | 1 |
Run 2 | 27 (5) | 14 | 5 | 8 | 0 |
Run 3 | 29 (2) | 10 | 10 | 9 (3 NR) | 0 |
Total | 78 (16) | 40 | 17 | 21 | 1 |
MDA‐MB‐231 | |||||
Run 1 | 22 (9) | 20 | 2 | 0 | 0 |
Run 2 | 27 (5) | 25 | 1 | 1 | 0 |
Run 3 | 29 (2) | 20 | 4 | 5 (2 NR) | 0 |
Total | 78 (16) | 65 | 7 | 6 | 0 |
Distribution of results from participating laboratories over three runs during 2004/05 as assessed by results from reference laboratories.
Results = number of participating laboratories submitting results. Of participating laboratories, 22 participated in run 1, 25 in run 2 and 29 in run 3; figures in brackets represent those participating laboratories which did not submit results for that run (e.g. in run 1, 22 laboratories submitted results and 9 did not).
Results were scored as either “appropriate” (falling within the range identified by the reference laboratories), “acceptable” (falling within 10% of the range identified by the reference laboratories), “inappropriate” (falling outside the range ±10% identified by reference laboratories but not resulting in a misdiagnosis), or “misdiagnosis” (falling outside the above range and resulting in a possible “misdiagnosis” of the case in question).
*BT20 was used in run 1 and MDA‐MB‐175 was used for runs 2 and 3.