. 2007 Jan 23;63(6):722–731. doi: 10.1111/j.1365-2125.2006.02835.x

Table 2.

Mean (CV) penetration index (PI) and deposition of radioactivity in the lungs, trachea/oesophagus, and oral cavity immediately after nebulization and subsequent mouth rinsing

Deposition of radioactivity
Radioaerosol Nebulizer PI Lungs (MBq 99mTc) Trachea/oesophagus (MBq 99mTc) Oral cavity(MBq 99mTc) Trachea/oesophagus andoral cavity (% of total)
99mTc-DTPA Pari 1.8 (24%) 7.5 (14%) 0.4 (52%) 0.4 (59%) 10
99mTc-DTPA UltraVent 2.4 (11%) 5.5 (26%) 0.1 (38%) < 0.1 n.a. < 4
99mTc-Nanocoll Pari 1.7 (21%) 12.9 (34%) 0.5 (87%) 0.7 (77%)  9

n.a. = not applicable.