Table 2.

Global and regional apparent uptake differences (using equation 1) between the three CT attenuation corrected cardiac uptakes. The cardiac uptake computed using the end-expiration CT (UpCTEXPIR) is the reference. Negative signs imply the value is lower than for UpCTEXPIR.

Regional Uptake Bias (%)UpCTEXPIR as reference UpCTMIDVOL UpCTINSPIR
Global −14.7 ± 8.2 −18.0 ± 11.0

Apex −8.3 ± 6.3 −16.8 ± 14.2

Apical slice
 Anterior −17.6 ± 17.5 −30.2 ± 23.6
 Inferior −2.9 ± 5.7 −7.9 ± 8.5
 Lateral −14.1± 12.4 −25.5 ± 17.6
 Septal −2.4 ± 5.7 −7.1 ± 8.3

Mid slice
 Anterior −14.8 ± 12.4 −26.7 ± 19.5
 Inferior −4.5 ± 7.2 −11.9 ± 9.8
 Lateral −14.7 ± 8.5 −29.0 ± 15.2
 Septal −3.1 ± 4.4 −8.1 ± 7.7

Basal slice
 Anterior −10.3 ± 7.2 −18.4 ± 13.5
 Inferior −6.7 ± 7.0 −13.2 ± 10.4
 Lateral −12.3 ± 7.8 −24.4 ± 13.6
 Septal −5.1 ± 4.9 −9.3 ± 7.2

All p<0.05