Skip to main content
. Author manuscript; available in PMC: 2007 Oct 18.
Published in final edited form as: IEEE Trans Med Imaging. 2003 Oct;22(10):1307–1318. doi: 10.1109/TMI.2003.818263

TABLE I.

Summary of Errors for Phantom Study

Two Views (°) Absolute Distance Error (mm)
RAO LAO Mean Min Max Standard Deviation
0 70 4.00 0.26 15.00 2.47
10 70 4.06 0.16 14.89 2.37
20 70 4.68 0.14 15.81 2.90
30 60 4.44 0.17 15.46 2.82
40 50 4.78 0.16 15.92 3.76
50 40 4.07 0.06 12.52 2.51

Four Views (°)
RAO 0, LAO 70; RAO 30, LAO 60 3.80 0.19 10.49 2.05