TABLE 2.

Calculated Z′ values as a measure of reproducibility of the assays with various strains

Strain Za
Dynamic rangeb
OD determination
Luminescence assay
OD determination
Luminescence assay
Growth plates
CV assay with pins
CV assay with plates
Blank Mutant control Growth plates CV assay with pins CV assay with plates
Blank Mutant control Blank Mutant control Blank Mutant control
PA14 0.68 <0.1 0.67 0.60 0.57 0.37 0.89 0.82 0.164 (0.007) 0.268 (0.016) 0.967 (0.061) 992,640 (36,261)
ZK2870 <0.1 0.53 0.80 0.77 0.77 0.70 0.92 0.89 0.085 (0.006) 0.503 (0.024) 1.543 (0.063) 2,006,927 (52,801)
PAO1 0.60 <0.1 0.77 0.75 0.47 0.32 0.86 0.82 0.127 (0.005) 0.330 (0.013) 0.920 (0.057) 1,715,298 (80,165)
PAO1 ΔfliC 0.68 NA <0.1 NA <0.1 NA 0.87 NA 0.165 (0.007) 0.110 (0.005) 0.471 (0.043) 225,330 (9,187)
PAK 0.60 <0.1 0.27 0.27 <0.1 <0.1 0.69 0.64 0.173 (0.007) 0.138 (0.008) 0.269 (0.024) 436,591 (40,870)
PAK ΔfliC 0.56 NA <0.1 NA <0.1 NA <0.1 NA 0.158 (0.006) 0.085 (0.005) 0.225 (0.023) 26,611 (8,569)
Blank NA NA NA NA NA NA NA NA 0.057 (0.004) 0.091 (0.005) 0.319 (0.031) 156 (77)
a

The mutant control used for the Z′ calculations for strains PA14, ZK2870, and PAO1 was PAO1 ΔfliC, whereas the mutant control used for the Z′ calculation for strain PAK was PAK ΔfliC. NA, calculation is not applicable.

b

The dynamic range data consist of raw values observed and demonstrate that luminescence had a range of 12,865 times over that for the blank control, whereas the dynamic range for the OD was a maximum of 5.5 times over that for the blank control. Standard deviations are indicated in parentheses.