Skip to main content
. 2007 Aug 3;73(19):6296–6298. doi: 10.1128/AEM.00325-07

TABLE 1.

Performance of modified Kinyoun acid-fast staining processed with filter water, tap water, and deionized water

Processing water and culture status No. of smears
%
Positive Negative Sensitivity Specificity PPVa NPVb
Filterc 44.3 98.6 83.3 91.5
    Positive 35 44
    Negative 7 476
Tap 51.9 87.6 40.6 91.8
    Positive 41 38
    Negative 60 423
Deionized 53.2 89.9 46.2 92.1
    Positive 42 37
    Negative 49 434
a

Proportion of specimens with positive culture results which are correctly detected by smear.

b

Proportion of specimens with negative culture results which are correctly detected by smear.

c

P values for sensitivity, specificity, PPV, and NPV in comparison with tap water are 0.17, <0.001, <0.001, and 0.45, respectively, and in comparison with deionized water are 0.13, <0.001, <0.001, and 0.36, respectively.