Table 4.
Needle electrode configuration | U/d = 1.15 V/cm | U/d = 1300 V/cm | ||||
Target tissue Ettmax(V/cm) |
Target tissue Ettmin(V/cm) |
Entire tissue Emax(V/cm) |
Target tissue Ettmax(V/cm) |
Target tissue Ettmin(V/cm) |
Entire tissue Emax(V/cm) |
|
2 (Fig. 4a) | 0.804 | 0.527 | 6.618 | 905.4 | 591.7 | 7450.3 |
4 (Fig. 4b) | 0.824 | 0.779 | 5.829 | 928.7 | 876.9 | 6562.5 |
6 (Fig. 4c) | 1.049 | 0.911 | 5.166 | 1180.9 | 1025.2 | 5816.6 |
6 (Fig. 4d) | 0.822 | 0.778 | 5.794 | 925.4 | 875.9 | 6523.1 |
6 (Fig. 4e) | 1.064 | 0.835 | 7.443 | 1197.9 | 940.1 | 8379.6 |
7 (Fig. 4f) | 0.21 | ~0 | 5.17 | 236.4 | 1.038 | 5820.8 |
7 (Fig. 4g) | 8.1 | 0.84 | 8.1 | 9119.2 | 945.7 | 9119.2 |
The results for models shown in Figs. 4a – 4e were calculated for applied voltage U = 1 V () and for models shown in Figs. 4f – 4g for applied voltage U = 0.575 V (d = l = 5 mm) giving U/d = 1.15 V/cm. Furthermore, by scaling the results we also calculated the parameters Ettmin, Ettmax and Emax for U/d = 1300 V/cm.