Table 6.
Needle electrode configuration | U/d = 1.15 V/cm | U/d = 1300 V/cm | ||||
Target tissue Ettmax(V/cm) |
Target tissue Ettmin(V/cm) |
Entire tissue Emax(V/cm) |
Target tissue Ettmax(V/cm) |
Target tissue Ettmin(V/cm) |
Entire tissue Emax(V/cm) |
|
2 (Fig. 6a) | 0.558 | 0.364 | 7.001 | 628.05 | 409.9 | 7887.1 |
4 (Fig. 6b) | 0.572 | 0.539 | 6.104 | 643.4 | 606.2 | 6872.5 |
6 (Fig. 6c) | 0.741 | 0.639 | 5.339 | 833.9 | 718.9 | 6011.2 |
Quantification of the electric field strength (E) for in-homogeneous models with needle electrode configurations which are defined in Fig. 2a-c. Conductivity of the target tissue is σtt = 0.4 S/m and the conductivity of the surrounding tissue σst = 0.2 S/m. Calculated minimal E (Ettmin) and maximal E (Ettmax) inside the target tissue and maximal E within the entire tissue Emax for different needle electrode configurations. The results for models shown in Figs. 6a-6c were calculated for applied voltage U = 1 V (). Furthermore, by scaling the results we also calculated the parameters Ettmin, Ettmax and Emax for U/d = 1300 V/cm.