Skip to main content
. Author manuscript; available in PMC: 2007 Dec 6.
Published in final edited form as: J Chromatogr B Analyt Technol Biomed Life Sci. 2007 Aug 23;858(1-2):302–306. doi: 10.1016/j.jchromb.2007.08.022

Table 2.

Assessment of accuracy and precision from quality control samplesa

Nominal
(ng/mL)
GM
(ng/mL)
S.D.
(ng/mL)
DEV
(%)
BRP
(%)
WRP
(%)
n
3.00 2.99 0.10 -0.49 1.72 2.64 20
200 190.70 4.76 -4.65 1.36 2.19 20
400 378.91 13.43 -5.27 1.68 3.22 20
2000 1876.8 75.90 -6.16 3.31 2.78 20

Abbreviations: GM, grand mean; S.D., standard deviation; DEV, percent deviation from nominal value; WRP, within-run precision; BRP, between-run precision; n, total number of replicate observations within the validation runs.

a

Five samples at each concentration were run on four different occasions.