Typical operating conditions of Pb sensors based on Hg-film electrodes and SAMMS-based electrodes.

Parameter Condition
Flow system FIA/ASV at Hg-film electrode
 Sample preparation a) Pb-spiked samples containing 10–50 vol% urine, 10 vol% bood, or 50 vol% saliva; b) 10-min incubation; c) acidified to 1 M HCl; and d) filtered through 5,000 NMWL cut-off ultrafilters
 Sample volume 60 μL per measurement
 Carrier 5 ppm Hg/0.5 M HCl
 Flow rate 1 μL/sec
 Preconcentration –1.0 V, 110 sec
 Detection by SWV Scan from –0.75 V to –0.45 V
 Electrode precondition 0.6 V, 90 sec in carrier
Batch measurements SAMMS-based electrodes
 SAMMS 5–10 wt% of SH-, AcPhos-, or IDAA-SAMMS
 Conductive matrix Cabon paste, graphite ink, or carbon nanotube paste
 Samples Pb in DI water, groundwater, or samples contained 25% rat urine
 Sample volume 8 mL waters or 0.5 mL of urine samples
 Preconcentration 2–5 min in stirred solutions at open circuit
 Electrolysis –0.8 V to –1 V for 60 sec in 0.3–0.5 M HNO3 or HCl
 Detection by SWV Scan from –0.7 V to –0.35 V
 Electrode cleaning Often not required

Abbreviations: AcPhos, acetamide phosphonic acid; DI, deionized; IDAA, iminodiacetic acid; SWV, square-wave stripping voltammetry.