. 2007 Oct 9;16(12):2159–2165. doi: 10.1007/s00586-007-0444-5

Table 1.

Hybrid testing results

Ranges of motion (degrees) T12–L1 L1–L2 L2–L3 L3–L4 L4–L5 L5–S1
A. Flexion/extension
Intact 5.82 ± 1.86 6.78 ± 2.02 5.90 ± 1.43 7.34 ± 1.41 8.26 ± 2.57 4.73 ± 3.40
StabilimaxNZ® 6.65 ± 1.88* 7.05 ± 2.04 6.34 ± 1.83 8.49 ± 1.81 5.44 ± 1.95* 4.97 ± 3.62
Simulated fusion 7.13 ± 2.11 7.53 ± 1.76 6.91 ± 2.04 9.79 ± 2.77 2.04 ± 1.20* 5.27 ± 4.88
B. Bilateral lateral bending
Intact 6.58 ± 2.22 7.21 ± 2.25 9.25 ± 1.61 8.32 ± 2.84 7.81 ± 2.99 5.31 ± 3.44
StabilimaxNZ® 7.04 ± 2.25 8.22 ± 3.09 10.00 ± 1.81 8.57 ± 2.68 4.85 ± 1.65* 5.91 ± 3.75
Simulated fusion 7.39 ± 2.36* 9.00 ± 3.59* 10.95 ± 1.98* 9.91 ± 3.11* 0.89 ± 0.34* 7.21 ± 4.93*
C. Bilateral torsion
Intact 3.11 ± 2.30 3.56 ± 2.48 4.02 ± 1.95 5.19 ± 2.31 4.12 ± 1.26 2.85 ± 1.52
StabilimaxNZ® 3.01 ± 2.28 3.7 ± 2.66 3.93 ± 1.84 5.38 ± 2.36 4.86 ± 1.46 2.73 ± 1.29
Simulated fusion 3.63 ± 2.61* 4.32 ± 2.89* 4.70 ± 2.36* 6.26 ± 2.58* 1.49 ± 1.09* 3.80 ± 1.71*

Average ± 1 SD of intervertebral ranges of motion (iROM) in degrees for the Intact, StabilimaxNZ®, and simulated fusion

*Significant (P < 0.05) difference with respect to the corresponding Intact values