Summary of effective orifice area (5 lpm, and 75 bpm).

Cardiac output [lpm] PVR [mm Hg/lpm] EOA [cm2]
SJM value 5.1±0.1 3.05±0.1 3.64
OTD value 5.2±0.1 2.9±0.1 2.45
Diode 5.1±0.1 3.0±0.1 1.24
Nozzle 5.1 ±0.2 3.0±0.1 1.13