Table 6.
ABI 3100 Genetic Analyzer Electrophoresis Conditions for the Common Platform Using the POP6 Polymer and a 36 cm Array and Data Collection Software v2.0
Parameter | Data collection |
---|---|
Shorter [5–70 (CGG)n] repeats | |
Running temperature | 60°C |
Cap fill volume | NA |
Current tolerance | NA |
Run current | NA |
Voltage tolerance | NA |
Pre-run voltage | 15 kV |
Pre-run time | 180 seconds |
Injection voltage | 10 kV |
Injection time | 1 second |
Run voltage | 15 kV |
Number of steps | 10 nK |
Voltage step interval | 60 seconds |
Data delay time | 300 seconds |
Run time | 3000 seconds |
Longer [71–230 (CGG)n] repeats | |
Running temperature | 60°C |
Cap fill volume | NA |
Current tolerance | NA |
Run current | NA |
Voltage tolerance | NA |
Pre-run voltage | 15 kV |
Pre-run time | 180 seconds |
Injection voltage | 15 kV |
Injection time | 22 seconds |
Run voltage | 15 kV |
Number of steps | 10 nK |
Voltage step interval | 60 seconds |
Data delay time | 300 seconds |
Run time | 5600 seconds |