Table 6.

ABI 3100 Genetic Analyzer Electrophoresis Conditions for the Common Platform Using the POP6 Polymer and a 36 cm Array and Data Collection Software v2.0

Parameter Data collection
Shorter [5–70 (CGG)n] repeats
 Running temperature 60°C
 Cap fill volume NA
 Current tolerance NA
 Run current NA
 Voltage tolerance NA
 Pre-run voltage 15 kV
 Pre-run time 180 seconds
 Injection voltage 10 kV
 Injection time 1 second
 Run voltage 15 kV
 Number of steps 10 nK
 Voltage step interval 60 seconds
 Data delay time 300 seconds
 Run time 3000 seconds
Longer [71–230 (CGG)n] repeats
 Running temperature 60°C
 Cap fill volume NA
 Current tolerance NA
 Run current NA
 Voltage tolerance NA
 Pre-run voltage 15 kV
 Pre-run time 180 seconds
 Injection voltage 15 kV
 Injection time 22 seconds
 Run voltage 15 kV
 Number of steps 10 nK
 Voltage step interval 60 seconds
 Data delay time 300 seconds
 Run time 5600 seconds