Table 3.

Results of dose comparisons for the benchmark case, obtained during the credentialing process. The RPC calculated doses that were 2 % to 3 % higher on average, with standard deviations of 4 % to 6 %.

MCD APDc APDs

HDR Average (RPC/ Inst) 1.029 1.032 1.022
Std. Dev. (%) 4.0 % 4.2 % 3.5%
LDR Average (RPC/ Inst) 1.020 1.022 1.011
Std. Dev. (%) 6.0% 6.0% 5.8%