. 2008 Jan;21(1):97–110. doi: 10.1128/CMR.00035-07

TABLE 4.

Performance characteristics of BinaxNOW Malaria kit for Plasmodium falciparum and P. vivaxa

Characteristic Plasmodium falciparum
Plasmodium vivax
% Sensitivity 95% confidence interval % Sensitivity 95% confidence interval
Detection of parasitemia level (per μl)
    >5,000 99.7 98-100 93.5 91-96
    1,000-5,000 99.2 96-100 81.0 76-85
    500-1,000 92.6 76-99 47.4 36-59
    100-500 89.2 75-97 23.6 17-31
    0-100 53.9 37-70 6.2 3-12
    Overall 95.3 93-97 68.9 66-72
        Specificity 94.2 93-95 99.8 99-100
Detection in venous vs finger stick samples 100 vs 98.8 96-100 vs 94-100 81.6 vs 80.6 74-87 vs 73-87
    Specificity 94.7 vs 90.4 93-96 vs 88-92 99.7 vs 99.5 99-100 vs 99-100
a

Data were obtained from the package insert and from selected references (28, 60).