Effects of collisional
quenchers on TMRM fluorescence near the S4 segment. (A,
left) The intensity ratio for constructs M356C W434F and
A359C W434F was computed by
dividing the fluorescence intensity after the application of 50
mM KI + 70 mM NMG-Mes by
the fluorescence intensity before
iodide application. Values <1.0
indicate a decrease in fluorescence after application. Fluorescence intensities were measured
at either −90 (black) or 0 (white)
mV, with error bars representing
SEM (M356C, n = 7; A359C, n =
4). The statistical likelihood of
no effect (P) was calculated as
the probability of obtaining these
statistics, given a true ratio of 1.
For all four measurements, P <
0.01. (right) Ratio of the iodide intensity ratio at −90 mV (R−90 mV)
to intensity ratio at 0 mV (R0 mV)
for constructs M356C W434F and
A359C W434F, with error bars =
SEM (M356C, n = 7; A359C, n =
4). Values >1 indicate that the
intensity ratio at −90 mV is
larger than the intensity ratio at 0 mV. M356C, P < 0.05; A359C, P = 0.13. (B, left) The intensity ratio for constructs M356C W434F and
A359C W434F was computed by dividing the fluorescence intensity after the application of 120 mM NMG-Mes in D2O by the fluorescence
intensity before D2O application. Values >1.0 indicate an increase in fluorescence after application. Fluorescence intensities were measured at either −90 (black) or 0 (white) mV, with error bars = SEM (M356C, n = 6; A359C, n = 6). For all four measurements, P < 0.01.
(right) Ratio of the D2O intensity ratio at −90 mV (R−90 mV) to intensity ratio at 0 mV (R0 mV) for constructs M356C W434F and A359C
W434F, with error bars = SEM (M356C, n = 6; A359C, n = 6). Values <1 indicate that the intensity ratio at −90 mV is smaller than the intensity ratio at 0 mV. For both measurements, P < 0.01.