Table 3.
Table 3 Percentage of Time That Manually Charted Settings Were Correct (Matched the MIB Values) and Average Number of Changes Recorded for Each Setting (HELP and MIB) per 24 Hours
| Setting | HELP Changes ∗ | MIB Changes† | Total Hours | Incorrect Hours | Percentage of Time Correct |
|---|---|---|---|---|---|
| Fio2 | 2.16 | 4.01 | 41,660 | 1,116.1 | 97.3 |
| PEEP | 0.85 | 2.33 | 41,660 | 412.2 | 99.0 |
| VT | 0.32 | 2.68 | 25,001 | 431.8 | 98.3 |
| Respiratory rate | 1.27 | 3.52 | 25,001 | 521.7 | 97.9 |
| Pressure support | 1.57 | 6.42 | 6,134 | 260.1 | 95.8 |
| Peak flow | 0.88 | 3.99 | 41,660 | 2,237 | 94.6 |
| Low VE alarm | 0.80 | 1.88 | 41,660 | 1,926 | 95.4 |
| High pressure alarm | 0.37 | 0.76 | 41,660 | 3,745 | 91.0 |
| High respiratory rate alarm | 0.69 | 2.19 | 41,660 | 7,977 | 80.9 |
| Low VT alarm | 0.90 | 3.12 | 41,660 | 10,032 | 75.9 |
∗ Average number of setting changes recorded in the computer charting application (HELP) per 24 hours.
† Average number of setting changes recorded by the MIB per 24 hours.
MIB = Medical Information Bus; Fio 2 = fraction of inspired oxygen; PEEP = positive end-expiratory pressure; VT = tidal volume; VE = minute ventilation.