Skip to main content
. Author manuscript; available in PMC: 2008 Feb 22.
Published in final edited form as: Diabetes Technol Ther. 2003;5(5):781–789. doi: 10.1089/152091503322526987

Table 4.

Comparison of Accuracy of Original and Modified Sensors

Original Sensors N=5,658 Paired Values Modified Sensors N=1,120 Paired Values P-value
Relative Absolute Difference Median (25th, 75th percentiles) 19% (9%, 36%) 11% (5%, 23%) < 0.001
Modified Clarke Error Grid Zone A percentage 58% 78% < 0.001
Modified Clarke Error Grid Zone A+B percentage 93% 98% <0.001
ISO Criteria Met According to Reference Glucose Level*
  ≤70 mg/dL 41% 48% 0.37
  71–120 mg/dL 45% 60% 0.07
  121–180 mg/dL 53% 70% 0.02
  181–240 mg/dL 60% 77% 0.004
  >240 mg/dL 58% 81% 0.007
  All Values 53% 72% < 0.001
Variation of Accuracy Among Sensors N=112 Sensors N=25 Sensors
 Sensors with Median RAD <10% 9% 36% 0.01
 Sensors with Median RAD <20% 54% 88% <0.001
 Sensors with Median RAD <30% 80% 100% <0.001
 Sensors with ≥80% of Values Meeting ISO Criteria 9% 24% 0.07
 Sensors with ≥60% of Values Meeting ISO Criteria 35% 80% < 0.001
*

ISO criteria: for reference glucose value ≤75 mg/dL, CGMS value ±15 mg/dL and for reference glucose value>75 mg/dL, CCGMS value within ±20%.

Excludes 3 original sensors with fewer than ten paired reference glucose values.