Skip to main content
. Author manuscript; available in PMC: 2008 Feb 25.
Published in final edited form as: Vaccine. 2007 Nov 20;26(2):193–200. doi: 10.1016/j.vaccine.2007.10.064

Table 2. Data of test samples determined to be acceptable a).

Sample dilution O.D. 405 %CV
≤ 1: 500 0.25 ≤OD< 0.3 <70% b)
0.3 ≤OD< 0.4 <50% b)
0.4 ≤OD< 0.5 <40% b)
0.5 ≤OD< 1.6 <30% b)
> 1: 500 0.25 ≤OD< 1.6 <30%
a)

First quality control of standard curve was tested for each ELISA plate, then the QC of unknown sample values were tested. Any data that did not match in the category above were not reliable.

b)

The values of %CV were decided based on the mean plus 2SD of samples in the indicated O.D. range (Fig. 6).