Skip to main content
. 2007 Dec;105:594–615.

TABLE 2.

CONCORDANCE, SENSITIVITY, AND SPECIFICITY OF THE COMPUTER PROGRAM (ROPtool) AND 3 INDIVIDUAL EXAMINERS IN DETERMINING THE PRESENCE OR ABSENCE OF PLUS DISEASE IN COMPARISON TO CONSENSUS OF 3 EXPERTS*

CONCORDANCE SENSITIVITY SPECIFICITY
ROPtool 175/185 (95%) 36/37 (97%) 139/148 (94%)
Examiner 1 172/185 (93%)
P = .5
24/37 (65%)
P < .001
148/148 (100%)
P = .01
Examiner 2 172/185 (93%)
P = .5
26/37 (70%)
P < .001
146/148 (99%)
P = .03
Examiner 3 168/185 (91%)
P = .1
30/37 (81%)
P = .02
138/148 (93%)
P = .8
*

P values are for the difference between each examiner and the computer program.