Table 1.

Specificity and limit of quantitation of 4-HPR in mouse plasma: lot-to-lot variation

Nominal conc. LLOQ (0.50 ng/mL)




Lot # Measured STDEV % CV % Accuracy
Lot 1 (n=3) 0.44 0.05 12.20 87.87
Lot 2 (n=3) 0.46 0.06 13.09 91.40
Lot 3 (n=3) 0.52 0.06 11.55 103.13
Lot 4 (n=3) 0.58 0.02 3.34 115.67
Lot 5 (n=3) 0.52 0.04 8.09 103.53
Lot 6 (n=3)
0.52
0.03
5.48
103.27
n 18
Mean 0.50
STDEV 0.05
% CV 0.10
% Accuracy 100.81