Skip to main content
. Author manuscript; available in PMC: 2009 Feb 1.
Published in final edited form as: J Chromatogr B Analyt Technol Biomed Life Sci. 2007 Dec 28;862(1-2):213–218. doi: 10.1016/j.jchromb.2007.12.013

Table 2.

Assessment of accuracy and precision from quality control samplesa

Drug Nominal (ng/mL) GM (ng/mL) SD (ng/mL) DEV (%) BRP (%) WRP (%) n
ABI-007 30 31.8 2.8 6.0% 8.1% 4.5% 19
ABI-007 150 159.1 8.4 6.1% 2.0% 5.0% 20
ABI-007 2,000 1,893.7 122.7 −5.3% 5.6% 4.2% 20
ABI-007 25,000 24,709.9 2,557.1 −1.2% 10.4% 4.1% 19
Paclitaxel 150 149.2 6.5 −0.5% 3.5% 2.9% 19
Unbound ABI-007 2,000 99.8 10.0 n/a 9.64% 5.14% 20

Abbreviations: GM, grand mean; SD, standard deviation; DEV, percent deviation from nominal value; BRP, between run precision; WRP, within run precision; n, total number of replicate observations within the validation runs.

*

Four or five samples at each concentration were analyzed on four different occasions.