Published in final edited form as: Ultrasound Med Biol. 2007 Oct 23;34(2):289–298. doi: 10.1016/j.ultrasmedbio.2007.07.015

Table 1.

Accuracy (95%) in °C for estimating temperature at 44°C in configurations tested with simulation, assuming normal distribution with computed standard deviation and slope (dCBE/dT) of CBE as function of temperature. Naq/Nlip denotes the ratio of aqueous to lipid scatterers assumed. Values marked baseline were used while varying other parameters, e.g., simulations for varying region size employed a 2:1 aqueous-to-lipid ratio and SNR of 19dB.

Mean + CBE Mean - CBE STD CBE

Scatterer Population Accuracy (±°C) Accuracy (±°C) Accuracy (±°C)

Naq/Nlip = 2 (baseline) 0.716 1.385 0.971
Naq/Nlip = 1 0.897 1.649 1.116
Random f(T) 0.612 2.222 1.227

Signal-to-Noise Ratio

13.5dB 0.917 1.839 1.175
19dB (baseline) 0.716 1.385 0.971
33dB 0.907 1.144 0.768

Region Size (cm2)

.5×.5 1.485 2.291 1.817
1×1 (baseline) 0.716 1.385 0.971
1×3 0.488 0.768 0.583