Table 1.
Settings for Positive End-Expiratory Pressure (PEEP), According to the Required Fraction of Inspired Oxygen (FiO2).*
FiO2 | PEEP |
---|---|
0.3 | 5 |
0.4 | 5–8 |
0.5 | 8–10 |
0.6 | 10 |
0.7 | 10–14 |
0.8 | 14 |
0.9 | 14–18 |
1.0 | 18–24 |
Settings are from the ARDSNet trial.19 The required FiO2 is the lowest value that maintains arterial oxyhemoglobin saturation above 90%. After the corresponding level of PEEP is selected, arterial oxyhemoglobin saturation and plateau airway pressure should be monitored in the patient.