TABLE 2.

Environmental stability testing of the anthrax lateral flow assay, using samples with ranges of concentrations of PA-specific IgG

Environmental treatment Valuea for indicated range of concn (μg/ml) of PA-specific IgG
0 4-10 11-15 20-50 150-180 >250
Thermal shock 0/20 3/12 7/8 20/20 20/20 20/20
High temp/high humidity 0/20 3/12 3/8 18/20 20/20 20/20
High temp/low humidity 0/20 2/12 3/8 20/20 20/20 20/20
Low temp/low humidity 0/20 7/12 8/8 20/20 20/20 20/20
Laboratory conditions 0/20 6/12 5/8 20/20 20/20 20/20
a

No. of positive reactions/no. of devices tested.