. 2008 Feb 13;46(4):1213–1219. doi: 10.1128/JCM.02036-07

TABLE 2.

Accuracy of bioluminescent antimicrobial susceptibility assay by species

Species No. of isolates No. of observationsa
% Correctb
Accuracy Tc
S R S R
Citrobacter freundii 2 7 5 100 100 100 −0.679
Citrobacter koseri 1 4 1 100 100 100 −0.762
Coagulase-negative Staphylococcus 2 6 3 100 67 83 −0.392
E. coli 48 227 59 98 95 97 −0.451
Enterobacter cloacae 3 9 8 78 100 89 −1.350
Enterococcus spp. 6 16 14 88 79 83 −0.316
Klebsiella oxytoca 2 10 2 100 100 100 −0.534
Klebsiella pneumoniae 14 55 19 87 68 78 −0.455
Proteus mirabilis 3 12 6 83 50 67 −0.278
Providentia stuartii 1 0 5
Staphylococcus aureus 3 12 6 92 33 63 −0.011
Serratia marcescens 1 3 3 100 67 83 −0.928
a

Number of observations containing antibiotic-susceptible (S) or antibiotic-resistant (R) organisms.

b

Calculated by dividing the number of specimens predicted correctly by the number tested.

c

T, threshold log RLU different from control result without antibiotics.