Skip to main content
. 2008 Feb 13;46(4):1517–1519. doi: 10.1128/JCM.02046-07

TABLE 1.

Comparison between Abbott RealTime HBV and Cobas TaqMan 48 HBV real-time PCR assays

Abbott RealTime assay result No. of results obtained by COBAS AmpliPrep/ Cobas TaqMan
<12 IU/ml, target detected 12-54 IU/ml Within dynamic range >8.04 log IU/ml Total
Negative 1 1
<10 IU/ml, target detected 14 1b 15
Within dynamic range 7a 1c 100 4d 112
>9.0 log IU/ml 0
Total no. 22 2 100 4 128
a

Abbott RealTime results ranged from 12 to 34 IU/ml.

b

Roche TaqMan, 34 IU/ml.

c

Roche TaqMan, 12 IU/ml; Abbott RealTime, 84 IU/ml.

d

Abbott RealTime results: 7.91, 8.15, 8.16, and 8.90 log IU/ml.