Table 1.
Simulated critical buckling load, Pcr, of CMC and PLA microneedles as a function of microneedle shape, length, and base width/diameter.
Conical (800 μm length)
|
Pyramidal (600 μm length)
|
Conical (600 μm length)
|
||||||
---|---|---|---|---|---|---|---|---|
Diameter* (μm) | Pcr (N)
|
Width* (μm) | Pcr (N)
|
Diameter* (μm) | Pcr (N)
|
|||
CMC | PLA | CMC | PLA | CMC | PLA | |||
50 | 0.0004 | 0.0020 | 50 | 0.0020 | 0.010 | 50 | 0.0007 | 0.0035 |
75 | 0.0019 | 0.0096 | 75 | 0.0087 | 0.044 | 75 | 0.0034 | 0.017 |
100 | 0.0061 | 0.031 | 100 | 0.026 | 0.13 | 100 | 0.011 | 0.055 |
200 | 0.10 | 0.51 | 200 | 0.36 | 1.8 | 200 | 0.18 | 0.91 |
300 | 0.5 | 2.6 | 300 | 1.8 | 8.9 | 300 | 0.94 | 4.7 |
400 | 1.7 | 8.4 | 400 | 5.5 | 28 | 400 | 3.0 | 15 |
600 | 8.6 | 43 | 500 | 13 | 67 | 500 | 7.3 | 37 |
800 | 27 | 137 | 600 | 27 | 137 | 600 | 15 | 76 |
Diameters and widths are reported at the base of the microneedles, which taper to a tip of approximately 25 μm diameter.