Published in final edited form as: Diabetes Technol Ther. 2003;5(5):791–800. doi: 10.1089/152091503322526996

Table 4.

Variation of Accuracy Among GW2B Sensor Pads (N= 171)*

% of Sensors
 Sensor Pads with Median RAD <10% 26%
 Sensor Pads with Median RAD <20% 68%
 Sensor Pads with Median RAD <30% 87%
 Sensor Pads with Median RAD <40% 97%
Values Meeting ISO Criteria
 Sensor Pads with ≥90% of Values Meeting ISO Criteria 8%
 Sensor Pads with ≥80% of Values Meeting ISO Criteria 27%
 Sensor Pads with ≥70% of Values Meeting ISO Criteria 40%
 Sensor Pads with ≥60% of Values Meeting ISO Criteria 57%
 Sensor Pads with ≥50% of Values Meeting ISO Criteria 70%
*

Includes only sensors with 10 or more paired reference glucose values.

ISO criteria: for reference s glucose value ≤75 mg/dL, GW2B value within ±15 mg/dL and for reference glucose value >75 mg/dL, GW2B value within ±20%.