Skip to main content
. 2008 Feb 26;466(5):1225–1231. doi: 10.1007/s11999-008-0169-8

Table 1.

Biomechanical testing results of an interlocking intramedullary nail and an inflatable nail

Biomechanical test Interlocking nail (GK) Expandable nail (FX) Paired difference (GK − FX) (p value) Median paired ratio (GK/FX)
Bending stiffness (N/mm)(n = 8) 222.7 (169.5–275.9) 280.8 (224.7–336.8) −58.1 (−103.5–12.6) (p = 0.02) 1.2
Rotational rigidity (Nm/degree)(n = 6) 2.72 (0.07–5.37) 2.80 (1.62–3.97) −0.08 (−3.22–3.07) (p = 0.95) 1.7
Play (degrees)(n = 6) 1.15 (−0.24–2.54) 0.38 (0.12–0.64) 0.77 (−0.77–2.32) (p = 0.26) 1.8
Maximum rotational strength (Nm)(n = 8) 26.1 (14.3–37.9) 12.6 (8.1–17.1) 13.49 (−0.08–27.05) (p = 0.05) 0.4

Values are expressed as means, with confidence intervals in parentheses.