Published in final edited form as: Invest Ophthalmol Vis Sci. 2008 Mar;49(3):954–960. doi: 10.1167/iovs.07-0493

TABLE 3.

Area under the Receiver Operating Characteristic Curve and Associated SE Are Presented for Each Parameter for Each Test (N = 166)

AUC SE Sensitivity/Specificity (%) Criteria for 80% Specificity Sensitivity/Specificity (%) Criteria for 90% Specificity
SAP-SITA
MD 0.680 0.041 46/80 1.98 38/90 2.49
 PSD 0.641 0.043 45/80 2.10 30/90 2.28
 TD <5% 0.668 0.042 46/81 13 27/91 20
 TD <1% 0.640 0.043 49/83 3 33/90 6
 PD <5% 0.645 0.043 51/79 9 31/90 13
 PD <1% 0.622 0.011 45/83 3 33/89 4
FDT N-30
 MD 0.660 0.042 44/80 −3.75 33/90 −4.74
PSD 0.733 0.038 56/80 4.91 40/90 5.72
 TD <5% 0.662 0.042 44/79 9 24/91 13
 TD <1% 0.689 0.041 54/78 2 40/90 4
 PD <5% 0.728 0.039 59/79 5 47/89 7
 PD <1% 0.690 0.041 56/73 1 42/90 2
FDT 24-2
 MD 0.763 0.037 55/80 −3.69 45/89 −5.09
 PSD 0.755 0.037 56/80 3.39 38/90 4.07
TD <5% 0.774 0.036 60/78 9 44/90 19
 TD <1% 0.761 0.037 64/78 2 46/90 6
 PD <5% 0.731 0.038 52/80 8 44/90 12
 PD <1% 0.756 0.037 62/75 3 47/90 5

Criteria for each parameter that yielded approximately 80% and 90% specificity are also presented, as are sensitivity at 80% and 90% specificity for each parameter of each test. Boldface text within the table indicates the best parameter for each test (MD for SAP, PSD for N-30, and TD < 5% for FDT 24-2). MD, mean deviation; PSD, pattern standard deviation; TD, total deviation; PD, pattern deviation.