Table.
Type of measurement | 95% limits of agreement half-widths (prism diopters) | 95% CI on limits of agreement (prism diopters) |
---|---|---|
Simultaneous prism and cover test at distance fixation | ± 6.3 | ± 3.7 |
Simultaneous prism and cover test at near fixation | ± 6.9 | ± 3.3 |
Alternate prism and cover test at distance fixation | ± 10.2 | ± 2.3 |
Alternate prism and cover test at near fixation | ± 9.2 | ± 2.5 |