Skip to main content
. Author manuscript; available in PMC: 2008 May 26.
Published in final edited form as: Magn Reson Med. 1995 Jul;34(1):92–98. doi: 10.1002/mrm.1910340114

Table 1.

Resistance Measurements between the Two Overlapped Coils on a Planar and a Cylindrical Frame

Phased array Sample Resonant frequency
(MHz)
R11
(Ω)
R22
(Ω)
R12
(Ω)
Cylindrical None 64.1 ± 0.1    5.3 ± 0.05    5.5 ± 0.05 12.9 ± 0.12
Human torso 63.9 ± 0.1  41.46 ± 0.42 55.45 ± 0.5 13.82 ± 0.13 
Planar None 63.9 ± 0.1 29.21 ± 0.3 19.09 ± 0.2 5.25 ± 0.05
Human torso 63.9 ± 0.1 56.72 ± 0.5 44.83 ± 0.4 15.83 ± 0.2