Published in final edited form as: Magn Reson Med. 1997 Sep;38(3):429–439. doi: 10.1002/mrm.1910380312

Table 1.

Summary for the Three Sources of Ghosting: Amplitude and Phase Discontinuities, and Time Delays. The first column is the amplitude of the m ± 1 ghosts normalized by the image intensity. The second column is the criteria of each source for reducing ghosts below the ghost to noise ratio, α. The third column is the minimum possible ghost to noise ratio using the reference method described in the text.

Ghost amplitude (m = ± 1) Discontinuity criteria Ghost to noise ratio
Amplitude |GA|=ΔA2nisin(π2ni) ΔA2αnisin(π2ni)SNR
Phase |Gϕ|=sin(Δϕ2)1nisin(π2ni) Δϕ2αnisin(π2ni)SNR αΔySyΔxSxNy2nisin(π2ni)
Time delay |Gs[nx]|=sin(2πsnxNx)1nisin(π2ni) tdΔxαnisin(π2ni)BWSxπSNR αΔySxΔxSy6Nynisin(π2ni)