Table B1. MR pulse sequence protocol for a breast MR examination in use for this study.

Sequence type Repetition time TR (ms) Echo time TE (MS) 1.5 T Flip angle (0) Matrix size Number of slices Field of view (FOV) (MM) Pixel size (mm x mm) Number of averages Acquisition time (min:s)
Precontrast, high-resolution 3D T1-weighted 8.9 4.2 35 512 × 384 60 340 0.66 × 0.89 1 2
Dynamic contrast enhanced, 3D T1-weighteda 2 sequences before contrast injection and 6 after 8.8 4.2 35 256 × 256b 480 340 1.33 × 1.33b 1 8 × 1.10
Postcontrast, high-resolution 3D T1-weighted with fat suppression 18.9 4.2 35 512 × 384 60 340 0.66 × 0.89 1 5
a

Contrast injection was 0.16 mmol kg−1 body weight with an injection speed of 3 ml s−1, given after the second precontrast dynamic sequence.

b

The matrix size is optimised to ensure that the acquisition time of each 3D image is 70 s.