Skip to main content
. Author manuscript; available in PMC: 2008 Jun 16.
Published in final edited form as: Nat Protoc. 2007;2(10):2354–2365. doi: 10.1038/nprot.2007.308

TABLE 3.

Troubleshooting table.

Step Problem Possible reason Solution
R5 labeling
7 Precipitation when spin label is added to the nucleic acid solution Nucleic acid (generally 4100 nt) crashed out of solution Reduce the amount of acetonitrile
Phase separation Poor mixing between the aqueous solution and the acetonitrile solution of nitroxide Constant and gentle vortexing during incubation
8 Broad elution profile without clearly distinguishable peaks Nucleic acid forms secondary structures that interfere with HPLC separation Purify using denaturing gel electrophoresis
Doublet in the HPLC traces, all elute earlier than the unlabeled species Separation of the phosphorothioate diastereomers in the labeled oligonucleotide Collect all peaks and confirm R5 labeling by mass spectrometry and continuous-wave-electron paramagnetic resonance (cw-EPR); All R5-labeled fractions may be combined in subsequent steps
Multiple peaks in HPLC traces, some of them elute at the same position as the unlabeled species Incomplete R5 labeling Verify the presence of phosphorothioate in the oligonucleotide by mass spectrometry, I2 cleavage and nuclease digestion
Increase incubation time at Step 7
Increase the amount of compound 1
DEER data acquisition
14 The temperature does not decrease Presence of an obstruction in the system Check that the needle valve which controls the flow of liquid helium is open
If the needle valve works well, disassemble the transfer line, warm up to room temperature and purge with dry air or nitrogen
In rare instances, the obstruction may be in the cryostat. If purging the transfer line does not solve the issue, warm up the entire system and purge it carefully before reassembly
15 The ‘dip’ cannot be located The system is over- or undercoupled Adjust the coupling while scanning the frequency to locate the ‘dip’
The system has automatically entered the pulse mode to protect the hardware Enter the FT Bridge window and use the Bridge/Standard Configuration tab to select the cw mode
16 The slight shift in frequency is not observed when the sample is inserted in the cryostat The capillary may not be positioned correctly in the cavity Remove the sample and insert it again
Ice forms on the capillary Long exposure of the frozen capillary to air Keep capillary immersed in liquid N2 until loading
Capillary stuck to the resonator Ice introduced into the resonator Warm up and purge the system
‘Dip’ disappears Ice introduced into the resonator Warm up and purge the system
Capillary shatters Capillary not properly sealed Prepare multiple sample capillaries
21 Echo signal is clipped Video gain value is too high Display the signal without averaging and set the video gain so that the echo is not clipped
SpecJet Display gives an error message One or more of the timings may be incorrect Verify that all the time intervals are set correctly
Refocused echo and Hahn echo are weak or cannot be detected Settings are not optimized properly Verify that center field and field position are set properly; Readjust settings
Sample not in the active volume Reinsert the sample
Spin concentration may be too low Collect a cw-EPR spectrum to verify spin concentration. Relabel as needed
Strong Hahn echo, but weak or non-existence of the refocused echo Sample has short-phase memory time Try to increase phase memory time by lowering temperature or changing solvent (D2O instead of H2O)
23 Only mono-exponential echo decay is observed Low concentration of interacting spin pairs Check and adjust labeling and assembling procedures to ensure sufficient presence of double-labeled sample
NASNOX-W
38 The NASNOX-W Applet fails to run Errors in the PDB coordinates; incorrect atom names in the PDB file Check structure correctness using a graphics program; check atoms names, particularly of phosphates
39 Corrupted structure in ‘data001lig.pdb’ Errors in writing the output structure due to variations in the PDB input file (e.g., modified nucleotides) Check the input PDB file format. However, in most such cases, distance information in the ‘data.add’ file is preserved