TABLE 3.
Troubleshooting table.
Step | Problem | Possible reason | Solution |
---|---|---|---|
R5 labeling | |||
7 | Precipitation when spin label is added to the nucleic acid solution | Nucleic acid (generally 4100 nt) crashed out of solution | Reduce the amount of acetonitrile |
Phase separation | Poor mixing between the aqueous solution and the acetonitrile solution of nitroxide | Constant and gentle vortexing during incubation | |
8 | Broad elution profile without clearly distinguishable peaks | Nucleic acid forms secondary structures that interfere with HPLC separation | Purify using denaturing gel electrophoresis |
Doublet in the HPLC traces, all elute earlier than the unlabeled species | Separation of the phosphorothioate diastereomers in the labeled oligonucleotide | Collect all peaks and confirm R5 labeling by mass spectrometry and continuous-wave-electron paramagnetic resonance (cw-EPR); All R5-labeled fractions may be combined in subsequent steps | |
Multiple peaks in HPLC traces, some of them elute at the same position as the unlabeled species | Incomplete R5 labeling | Verify the presence of phosphorothioate in the oligonucleotide by mass spectrometry, I2 cleavage and nuclease digestion | |
Increase incubation time at Step 7 | |||
Increase the amount of compound 1 | |||
DEER data acquisition | |||
14 | The temperature does not decrease | Presence of an obstruction in the system | Check that the needle valve which controls the flow of liquid helium is open |
If the needle valve works well, disassemble the transfer line, warm up to room temperature and purge with dry air or nitrogen | |||
In rare instances, the obstruction may be in the cryostat. If purging the transfer line does not solve the issue, warm up the entire system and purge it carefully before reassembly | |||
15 | The ‘dip’ cannot be located | The system is over- or undercoupled | Adjust the coupling while scanning the frequency to locate the ‘dip’ |
The system has automatically entered the pulse mode to protect the hardware | Enter the FT Bridge window and use the Bridge/Standard Configuration tab to select the cw mode | ||
16 | The slight shift in frequency is not observed when the sample is inserted in the cryostat | The capillary may not be positioned correctly in the cavity | Remove the sample and insert it again |
Ice forms on the capillary | Long exposure of the frozen capillary to air | Keep capillary immersed in liquid N2 until loading | |
Capillary stuck to the resonator | Ice introduced into the resonator | Warm up and purge the system | |
‘Dip’ disappears | Ice introduced into the resonator | Warm up and purge the system | |
Capillary shatters | Capillary not properly sealed | Prepare multiple sample capillaries | |
21 | Echo signal is clipped | Video gain value is too high | Display the signal without averaging and set the video gain so that the echo is not clipped |
SpecJet Display gives an error message | One or more of the timings may be incorrect | Verify that all the time intervals are set correctly | |
Refocused echo and Hahn echo are weak or cannot be detected | Settings are not optimized properly | Verify that center field and field position are set properly; Readjust settings | |
Sample not in the active volume | Reinsert the sample | ||
Spin concentration may be too low | Collect a cw-EPR spectrum to verify spin concentration. Relabel as needed | ||
Strong Hahn echo, but weak or non-existence of the refocused echo | Sample has short-phase memory time | Try to increase phase memory time by lowering temperature or changing solvent (D2O instead of H2O) | |
23 | Only mono-exponential echo decay is observed | Low concentration of interacting spin pairs | Check and adjust labeling and assembling procedures to ensure sufficient presence of double-labeled sample |
NASNOX-W | |||
38 | The NASNOX-W Applet fails to run | Errors in the PDB coordinates; incorrect atom names in the PDB file | Check structure correctness using a graphics program; check atoms names, particularly of phosphates |
39 | Corrupted structure in ‘data001lig.pdb’ | Errors in writing the output structure due to variations in the PDB input file (e.g., modified nucleotides) | Check the input PDB file format. However, in most such cases, distance information in the ‘data.add’ file is preserved |