Table 3. Detailed statistics of Linco-plex assay recoveries (two independent experiments).

Cytokines IL-1α IL-1β IL-1rα (−) IL-2 IL-4 IL-5 IL-6 IL-7
Number of positive reads 6 7 4 7 7 7 5 7
Minimum 61.44 70.7 35.02 77.09 70.92 75.23 55.12 64.04
Median 87.88 93.63 82.98 87.9 92.48 112.8 78.96 77.5
Maximum 99.01 123.2 180.5 102.9 314.4 177 110.2 120.9
Mean 85.43 94.49 95.37 87.58 121.1 118.6 81.48 84.32
Std. Deviation 12.92 18.51 61.23 8.522 86.17 34.65 22.5 19.67
Std. Error 5.275 6.997 30.62 3.221 32.57 13.09 10.06 7.433
Coefficient of variation 15.13% 19.59% 64.21% 9.73% 71.15% 29.22% 27.62% 23.32%

(−) Cytokines not included in the second test.

Unstimulated whole blood culture supernatant samples were spiked at different concentrations with recombinant cytokines. Samples were analysed on the Bio-plex system instrument and the recovery of each cytokine assessed after subtraction of the endogenous levels of cytokines.