TABLE 2.

Sensitivity and specificity analysis of chromID VRE and EVA for VRE at 48 h

Medium No. of resultsa
Sensitivity
Specificity
TP FN % 95% CI % 95% CI
chromID VRE 32 1 96.9 84.3-99.5 99.4 98.1-99.8
EVA 31 2 93.9 80.1-98.4 83.4 79.1-86.6
a

A true positive (TP) is defined as a blue-green or purple colony on chromID VRE or a brown to black colony on EVA that was identified as VRE by VITEK 2 and confirmed by PCR. A false negative (FN) is defined as an isolate that was confirmed as a VRE on one medium but did not grow on the other medium.