Skip to main content
. 2008 Apr 18;95(5):2583–2598. doi: 10.1529/biophysj.107.118620

TABLE 1.

Optimal parameters of microscopy for mitochondrial swelling measurement with the TR technique

Important parameters Additional parameters
Fluorophore Large molecular weight, organelle-specific, bright λ is less important, can be a probe
Focus z-stacking and mean intensity projection Step size = axial resolution ∼0.8 μm
Optics NA > 0.8, but higher is better; if confocal, pinhole > 1.5 resel Wide-field (or confocal)
Detector 0.1–0.14 μm/pixel resolution ∼1000 photo e/pixel (summed for the z-stack) Use region averaging