TABLE 1.
Microscopy, Ziehl-Neelsen result | No. with Cryptodiag resulta:
|
|||
---|---|---|---|---|
Negative | Gray zone | Positive | Total | |
Negative | 114 | 2 | 2 | 118 |
Positive | 0 | 1 | 32 | 33 |
Total | 114 | 3 | 34 | 151 |
Positive, OD ≥ 0.4; gray zone, 0.3 ≤ OD < 0.4; negative, OD < 0.3. Cryptodiag sensitivity and specificity were as follows: for threshold of OD ≥ 0.4, sensitivity, 97.0% (32/33), and specificity, 98.3% (116/118); for threshold of OD ≥ 0.3 (including gray zone), sensitivity, 100% (33/33), and specificity, 96.6% (114/118).