Table 1.

Comparison of the Current Density at Different Decouplers as Increased Noise Is Apparent at the Working Electrode

decoupler type limiting separation current (μA) noise (pA) current density (A/cm2)
24 hole, laser etched 30 1.4 0.255
10 hole, laser etched 5 10 0.102
Fracture 5 1 0.212
300 μm, cast Nafion tube 20 0.5 0.0212
500 μm, cast Nafion tube 25 0.5 0.0283
1000 μm, cast Nafion tube 30 0.5 0.0127