Skip to main content
. Author manuscript; available in PMC: 2008 Sep 4.
Published in final edited form as: BJU Int. 2007 Feb 19;99(5):1134–1142. doi: 10.1111/j.1464-410X.2006.06736.x

Figure 7.

Figure 7

Photodiode measures of inter-pulse delay for a set of Duet electrodes fired in synchronous mode. Early in the life of the electrodes, firing was nearly simultaneous and the majority of paired pulses were separated by a delay of ≤2 μs (dashed line). As the electrodes aged, the inter-pulse delay increased, such that during the last 1400 paired pulses only ~27% were simultaneous (≤2 μs). Solid line tracks mean values for inter-pulse delay at 100 shot intervals.